MRO Magazine

Multichannel pipette calibration balance

October 1, 2023 | By MRO Staff

Photo: Mettler Toledo

Photo: Mettler Toledo

Mettler Toledo XPR105MCP has readability of 10 μg and repeatability of 15 μg, the system fulfills requirements of ISO 8655 for multichannel pipettes with dis-pensing volumes as small as 0.5 μL.

Results of a single dispense of up to 12 channels are calculated in under one minute. XPR105MCP tests each pipette channel individually on the same load cell to ensure equal measurement performance.

The 360 measurements needed for a full calibration (12 channels, 3×10 repeti-tions) can be completed in 25 minutes. The system can also be connected to a second XPR105MCP for increased throughput.

Software stores result in a secure database, informs you when pipette tests are due, and produces customized reports if required.